SUMMARY
The current study is a
comparative evaluation of the performance of collimators commonly used in SPECT. The
results show the difference in tomographic spatial resolution using a parallel hole high
resolution collimator and the parallel hole general purpose collimator and how the
distance between the collimator surface and the phantom influences the quality of the
image.
Key words: SPECT
systems, Jaszczack phantom, resolution, uniformity.
INTRODUCTION
The purpose of the current
study is a comparative evaluation of tomographic spatial resolution of two collimators
commonly used in SPECTS systems . This study was performed using a Jaszczak phantom ,
which is a cylindrical acrylic phantom with inserts of rods and spheres representing
"cold" areas, and a region of uniformity(1). The phantom study is
able to show the performance of the single photon emission computed tomography system
(SPECT) regarding spacial resolution and uniformity. The tomographic resolution is
determined through the observation of the cold rods section . The section without inserts,
filled with water and 851 MBq of Tc99m (TcO4_) is used as a
reference only for the visual uniformity analysis. The test with the phantom must be part
of a quality control routine program in SPECT systems, which enables the user to evaluate
the performance of the whole system every six months.
MATERIAL AND METHODS
For this study an acrylic
cylindric Jaszczak (1) phantom, with insert of spheres measuring 9.5, 12.7,
15.9, 19.1, 25.4 and 31.0 mm diameter and insert of rods measuring 4.8, 6.4, 7.9, 9.5,
11.1 and 12.7 mm diameter was used. The phantom was filled with water and 851 MBq activity
of Tc99m(TcO4_ ) were uniformly distributed. The images were
acquired with a single-head GE Starcam 3200 AC/T gamma camera having 3.1% integral
uniformity, 2.3% differential uniformity in the CFOV, 11.07% energy resolution and center
of rotation deviation of 0.4 mm. The photopeak was set at 140 keV with a 20% centered
energy window .
Image Acquisition:
One hundred and twenty eight
frames in 128 x 128 matrix were acquired with 1000 kc per frame in word mode and clockwise
circular 360o orbit. The collimators used were the parallel holes, Low Energy
High Resolution (LEHR) and Low Energy General Purpose (LEGP). The radius of rotation were
185 and 150 mm (a head holder was used in order to place the detector as close as possible
to the phantom). A bubble level was used to ensure that the phantom was levelled. It was
also adjusted so that the phantom would lie parallel the axis of rotation of the gamma
camera. The height of the scan table was adjusted so that the distance from the collimator
to the middle of the phantom was the same, with the detector at 0o and at 180o
.
Image Processing:
The images were processed using
Hanning filter with cut-off frequency of 1.6 cycles.cm-1, slice thickness of 5
pixels in the spheres section , 25 pixels at the rods section and 2 pixels at the uniform
section. A 100 million uniformity correction map and an attenuation coefficient of 0,11 -1
cm correction (2) were applied.
RESULTS:
1. Rods section
2. Spheres section
3. Uniform section
Absence of ring artifacts and uniform
radioactivity distribution.
Figure1
( click = zoom)
Transaxial slices of rods and spheres sections.
The images were obtained with LEGP collimator with 150 and 185 mm radius of rotation.
Figure 2
( click = zoom)
Transaxial slices of rods
and spheres sections. The images were obtained with LEHR collimator with 150 and 185 mm
radius of rotation.
Figure 3
( click = zoom)
Transaxial slices of the uniform
section.
DISCUSSION
It is well known that the
hight resolution collimators increase the ability to visualize small lesions and structure
in a lesion provided that count density is satisfactory, the best will be the detection
ability of the spECT system. The results show that the LEHR collimator offers the best
resolution when compared with the LEGP collimator, with a difference of 1.6 mm between
them. The best resolution was obtained with the LEHR collimator at the smallest radius of
rotation used in this study: 150 mm.
Ring artifacts were not seen at
the reconstructed images due to the application of an uniformity correction map with a
large total count (100 million kc).
CONCLUSION
The
"cost-effectiveness" of choosing collimators should be evaluated before the
decision for better resolution or better sensitivity for each particular study.
This study showed that the LEHR
collimator can contribute effectively to a better resolution than the LEGP collimator.
Although this conclusion is already available in the medical literature, mathematical data
were obtained for the equipment GE Starcam 3200 AC/T. These data allow other users to have
a better evaluation in comparison to other equipment commonly used in Nuclear Medicine.
REFERENCES
1. Data spectrums SPECT users
manual; Data Spectrum Corporation; Aug 1996.
2. OConnor, M.K.: The Mayo Clinic Manual
of Nuclear Medicine; 1996.
3. Muellehener et al Physics in Medicine and
Biology 30:163, 1985
4. Madsen et al. Physics in Medicine and Biology
37: 1625, 1992